NSN 6525-01-207-0824

Part Details | DENTAL X-RAY APPARATUS

6525-01-207-0824

Alternate Parts: G336, G-336, 6525-01-207-0824, 01-207-0824, 6525012070824, 012070824

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
65APR 27, 198501-207-082406317 ( X-RAY APPARATUS, DENTAL )
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6525-01-207-0824
Part Number Cage Code Manufacturer
G-33664616DEFENSE MEDICAL STANDARDIZATIONBOARD
Technical Data | NSN 6525-01-207-0824
Characteristic Specifications
DESIGN TYPEJR ANY ACCEPTABLE
TUBE TYPE SINGLE FOCUS
TUBE QUANTITY1
TUBE COOLING METHOD OIL
TUBE MOUNTING TYPE EXTENSION BRACKET
CONTROL TYPE CONSOLE
MOUNTING TYPE PORTABLE
OPERATION METHOD MANUAL
RADIOGRAPHIC TIMER TYPE HAND
TRANSFORMER RECTIFICATION METHOD SELF-RECTIFYING
RECTIFICATION TYPE HALF-WAVE
TRANSFORMER OUTPUT CAPACITY7 MA AT 70 KV PEAK
VOLTAGE IN VOLTSB110.0$$B220.0
CURRENT RATING IN AMPS10.0
FREQUENCY IN HERTZB50.0$B60.0
PHASE SINGLE
SPECIAL FEATURESTHE X-RAY UNIT AND COLLAPSIBLE PATIENT SEAT SHALL BE IN RUGGGED METAL CONTAINER,MADE OF 6 HANDLES. THE HANDLES SHALL BE 2 PER SIDE, ON OPPOSITE SIEDS, AND ONE PEER END PANEL, ON OPPOSITE ENDS. CONTIANER WILL HAVE PERMANENTLY ATTACHED GASKET SEAL (BUTYL RUBBER) RESIST ENTRY FROM RAIN AND DUST. CASE WILL BE CUSHIONED BY HIGH-DENSITY UNICELLULAR POLYETHYLENE FOAM. CASE WILL CONTAIN A PRESSURE RELIEF VALVE. UNIT MOUNTED ON TRIPOD IN THE CARRYING CASE. LEVELING METHOD SHALL BE EMPLOYED. THE CASE LEVELING BRACKET RECEPTACLES WILL BE REINFORCED TO PROTECT AGAINST TRANSPORT DAMAGES.
PART NAME ASSIGNED BY CONTROLLING AGENCYX-RAY APPARATUS FIELD DENTAL: