NSN 6665-01-370-0545
Part Details | RADIAC CALIBRATOR SET
6665-01-370-0545 A fixed number of components and/or items, not all having the same basic name, which are required for determination, by comparison, of the radioactivity detecting and measuring accuracy of radiac equipment, such as RADIAC SET. May or may not include comparison of computing accuracy.
Alternate Parts: PM1, 6665-01-370-0545, 01-370-0545, 6665013700545, 013700545
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|---|---|---|
66 | JAN 12, 1993 | 01-370-0545 | 00396 ( CALIBRATOR SET, RADIAC ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6665-01-370-0545
Part Number | Cage Code | Manufacturer |
---|---|---|
PM1 | 30653 | AIR FORCE METROLOGY AND CALIBRATION |
Technical Data | NSN 6665-01-370-0545
Characteristic | Specifications |
---|---|
COMPONENT QUANTITY | 6 |
SUPPLY ITEMS AND QUANTITIES | 6130-01-174-0653 POWER SUPPLY 1; 6625-01-251-5909 MULTIMETER 1; DIGITALTHERMOMETER 1; DOSIMETER 1; ELECTRONICBAROMETER 01; CAVITY IONIZATION CHAMBER 4; COAXIAL TERMINATION 1; LASER MEASUREMENT SYSTEM 1; ELECTROMETER 1 |
ACCESSORY COMPONENT QUANTITY | A1$$A1$$A1$$A1$$A1$$A1$$A1$$A1$$A1$$A1 |
ACCESSORY CONTROLLING AGENCY | KEITHLEY INSTRUMENTS INC; KITRICKMANAGEMENT CO; RUSKA INSTRUMENT CO; EXRADIN INC; EXRADIN INC; CHESAPEAKE LASER SYSTEMS INC |
SPECIAL FEATURES | CHAMBERS: A3 3/8 IN RADIUS,3.6 VOL CC,0.0 R/NC CAL FACTOR; A4 - 3/4 IN RAD,30 CC VOLUME,0.09 R/NC CAL; A51-1/8 IN RADIUS,100 CC VOL,0.03 R/NCCAL; A6 - 2-1/4 IN. RAD,800 CC VOL; 0.004 R/NC CAL; LASER: 1-6 METERS/SEC VELOCITY; 2 G'S ACCELERATION; DISTANCE .17-30 METERS; ELEVATAION PORM 60 DEG;ELECTROMETER: 200 MV-200 V RANGE; 10 UV-10 MV RESOLUTION; POWER SUPPLY: 0-3 KV DC OUTPUT; THERMOMETER: 0.1 DEG RESOLUTION; DISPLAYS DEGREES F. OR C.; BAROMETER: 0.5-19 PSIA,3.5-130 KPAA |
FUNCTIONAL DESCRIPTION | CALIBRATES GAMMA RADIATION SOURCES (STANDARDS) MODEL 81-10 ON SITE; ELECTROMETER USED TO MEASURE CHARGE ACCUMULATED BY IONIZATION CHAMBER; BAROMETER AND THERMOMETER MEASURE PRESSURE ANDTEMPERATURE AT POINT OF MEASUREMENT; LASER MEASUREMENT SYSTEM MEASURES DISTANCE FROM RADIATION SOURCE TO DETECTOR AND USED TO CALIBRATE RADIATION TRACK SYSTEM |